Installation

Make sure that the temperature at the site is within the
operating temperature range that is specified for the sampler.

Make sure that the intake tube is as short as possible.

Maximize the vertical slope of the intake tube from the sampler
to the sample source so that the tube drains completely. This
prevents cross-contamination of samples and freezing the tube.

Note: See

Specifications on page 5

for limitations on transport velocity

and maximum vertical lift.

Note: If site conditions do not permit the intake tube to have a
downward slope, or if the line is pressurized, disable the liquid sensor
and calibrate the sample volume (

section 4.8.2 on page 46

).

Install the strainer in the middle of the sample stream (not near
the surface or bottom) to make sure that a representative
sample is collected.

Figure 6 Mounting configuration

1

Strainer

4

Vertical lift

2

Intake tubing

5

Mounting surface

3

Suspension harness