45

Sampler operation

4.8.1 Volume calibration using the liquid sensor

Use the volume calibration with the liquid sensor enabled to adjust
sample volumes slightly.

Procedure:

1. Make sure the liquid sensor is enabled by selecting

PROGRAM

SETUP

>

MODIFY SELECTED

>

LIQUID SENSOR

>

ENABLE

.

2. Press

VOLUME CALIBRATION

and select

CALIBRATION

.

3. Remove the outlet pump tube from the fitting on the sampler

and place the tube into a graduated cylinder. Place the intake
tube into the sample source or tap water.

Note: Use the sample source for best accuracy.

4. Select

START

.

5. The pump will purge the intake tube and then collect the

sample volume that is specified in the sampling program. The
pump will then purge the intake tube.

To halt the calibration at any time, press the

STOP

key.

6. When the sample has been collected, select

DONE

. Compare

the volume collected in the graduated cylinder with the sample
volume that is entered in the program setup menu.

Select

REPEAT

to repeat the volume collection if necessary.

7. If the volume that was collected is different from the sample

volume in the current program, enter the volume that was
actually collected. Press

ENTER

. The sensor is then adjusted to

accurately measure all programmed volumes.

To verify sample volume, collect a grab sample (

section 4.8.3

on page 47

).

8. Select

START

to start the sampling program or

CANCEL

to exit

the volume calibration menu.

4.8.1.1 Reset the calibration

Use the reset cal option to restore the volume calibration to the
default factory settings.

Procedure:

1. Make sure the liquid sensor is enabled by selecting

PROGRAM

SETUP

>

MODIFY SELECTED

>

LIQUID SENSOR

>

ENABLE

.

2. To undo the calibration, press

VOLUME CALIBRATION

and

select

RESET CAL

.

3. The calibration will be reset to the default calibration and the

message “value has been reset to 0” will be shown.