4 sample history, 5 event log, 4 sample history 4.7.5 event log – Hach-Lange Sigma SD900 Portable User Manual

Page 45: Sampler operation 4.7.4 sample history

Advertising
background image

43

Sampler operation

4.7.4 Sample history

Sample history displays the sample number, bottle number, result
and time/date for each sample. The following results can be shown:

Success—the sample was drawn successfully

Bottle full—the full bottle shut-off was activated

Rinse error—an error occurred during the rinse cycle

User abort—a user pressed the

STOP

key to end the program

Arm faulty—the distributor arm did not move correctly

Pump fault—a fault occurred when the pump was operating

Purge fail—an error occurred during the purge cycle

Sample time out—liquid not detected within the timeout period

Power fail—a power failure occurred during sampling

If a program is running or has been halted, the sample history is
displayed for the currently running sample program. If the program
has been completed, the sample history is displayed for the most
recently completed sampling program. The sample history is
automatically erased when a new program is started.

Procedure:

1. Select

DIAGNOSTICS

>

SAMPLE HISTORY

from the main menu

to view the sample history.

2. The sample history will list the sample numbers, bottle numbers

and brief result for each sample. Select a sample from the list
and press ENTER to view complete details.

3. The complete details for the selected sample will be shown.

4.7.5 Event log

The event log records information on the following events:

Power on—power was connected to the controller

Power fail

—power was intentionally or unintentionally

disconnected from the controller without first turning power off

Firmware update—a new version of firmware was installed

Pump fault—a fault occurred when the pump was operating

Arm faulty—the distributor arm did not move correctly

Low memory battery—the internal battery should be replaced

User on—a user turned the power on using the power key

User off—a user turned the power off using the power key

Program start—the sample program was started

Program resume—a program was resumed from the halt state

Program halt—a program was halted

Program complete—a program was completed

Grab sample—a grab sample was taken

Change tube—the maximum pump cycle counts have expired