Sampler operation

4.8.3 Verify sample volume

To check the sample size after calibration use the Manual
Operation key to take a grab sample. Do not go back into
calibration to check the volume since the volume compensation is
reset to zero at the start of a calibration.

Procedure:

1. Press the

MANUAL OPERATION

key. Select

GRAB SAMPLE

.

2. Remove the outlet pump tube from the fitting on the sampler

and place the tube into a graduated cylinder. Place the intake
tube into the sample source or tap water.

3. If the liquid sensor is enabled, enter the volume that needs to

be verified. If the liquid sensor is disabled, enter the volume
specified in the program.

4. Press

ENTER

. The pump cycle will start (purge-sample-purge).

5. Compare the volume that was collected in the graduated

cylinder to the volume specified in step

3

. If the volume is

acceptable, the sampler is calibrated correctly. If the volume is
incorrect, repeat the volume calibration.

4.9 Controller settings

Use the system setup menu to change the following parameters:

Date and time

Communication

Sampler base

Liquid sensor calibration

Display contrast

Password

Pump tube replacement reminder

Procedure:

1. Select

SYSTEM SETUP

from the main menu.

2. Select one of the available parameters, for example time/date

setup. Refer to

Table 2 on page 35

for an overview of the

system setup menu.

3. Make selections or change values with the arrow keys. Use the

LEFT AND RIGHT

ARROW

keys to move the cursor left or right.

Use the

UP AND DOWN

ARROW

keys to change the value.

4. Select another parameter from the system setup menu, or

press

BACK

to return to the main menu.