Maintenance – Hach-Lange Sigma SD900 Portable User Manual

Page 65

Advertising
background image

63

Maintenance

Procedure:

1. Unscrew and remove the desiccant holder from the controller

(

Figure 26

).

2. Remove the plug from the end of the desiccant tube (

Figure 27

)

and discard the spent desiccant.

3. Fill the desiccant tube with fresh desiccant.

Note: Fresh desiccant should be yellow.

4. Replace the plug.

5. Apply grease to the O-ring (

Figure 27

).

6. Replace the desiccant tube in the controller.

Figure 26 Side view of controller—pump and desiccant

1

Desiccant window

2

Peristaltic pump

Figure 27 Desiccant holder

1

Plug

3

O-ring

2

Desiccant tube