Specifications – Hach-Lange Sigma SD900 Portable User Manual

Page 9

Advertising
background image

7

Specifications

Dimensions

10-

3

/

8

in. (26.4 cm) L x 11-½ in. (29.2 cm) W x 6-¾ in. (17.1 cm) H

Programming features

Password protection

6-character; protect changes to program and system settings

Multiple programs

Stores up to three sampling programs

Cascade programs

Two samplers used in combination. The second sampler is initiated after the first
sampler completes the program.

Synchronized sampling

Ability to simultaneously take two samples with input from a single flow meter

Sample volume

Programmed in 10-mL increments from 100 to 10,000 mL

Air purge

Air purged automatically before and after each sample; duration automatically
compensates for varying intake line lengths.

Intake rinse option

Option to rinse intake line with source liquid prior to each sample, 1 to 3 rinses.

Sample distribution

Composite, samples per bottle or bottles per sample.

Set point sampling

Ability to start and/or stop a sample program based on an external trigger.

User start/stop times

Up to 12 user-defined start/stop times/dates, with option to restart at position 1.

Storm water program

Ability to run time-based, first flush program in parallel with main sample program.

Current status

Display parameters relevant to main or storm water programs.

Units of measure

Volume: gallons or mL; length: feet (ft) or cm

Sample retries

Option to repeat sample collection cycle from 1 to 3 times if sample not obtained
on initial attempt.

Manual grab sample

Ability to manually deliver a grab sample to a specific bottle location

Run modes

Continuous or non-continuous with user-entered number of samples.

Time pacing

Uniform or variable time intervals.

Flow pacing

Uniform or variable flow intervals.

Auxiliary connector

Power to Sigma 9XX, SD900, flow pulse input, external inhibit, special output,
bottle number output and program complete output.

Program delay

Two formats: 1) 1–9,999 flow pulses (in one unit increments);
2) Programmable start time/date

Communication

Firmware updates

Ability to perform field upgrades using Sample View software

Serial interface

RS232 compatible; allows on-site collection of stored data including event log and
sample history. Ability to configure remotely. Supports Modbus for SCADA
connectivity.