3 rotor removal and cleaning, Maintenance 6.3.3 rotor removal and cleaning – Hach-Lange Sigma SD900 Portable User Manual
Page 62
60
Maintenance
6.3.3 Rotor removal and cleaning
Remove the pump rotor to thoroughly clean the silicone residue
from the pump tube.
Procedure:
1. Remove the pump tube (see
2. Open the rubber flap on the end of the rotor and remove the
screw with a flat-blade screwdriver (
3. Remove the rotor. A collet connects the rotor to the pump shaft
and can come off when the rotor is removed (
).
4. Clean the rotor, pump tube tracks and pump housing with a
mild detergent.
5. If the collect was removed in step
, align the recess in the
collet with the tang on the shaft and push the collet on the shaft
(
6. Push the rotor on the shaft and secure with the screw. Do not
overtighten the screw. Close the rubber flap on the rotor.
7. Install the pump tube (
).
Figure 23 Pump tube installed
1
Pump tube to sampler tube fitting
2
Pump tube to liquid sensor