Maintenance

Installation procedure:

1. Connect one end of the tubing to the liquid sensor fitting. Press

the tubing into the pump tube guide (

Figure 22

).

2. Press the tubing into the upper pump tube track. Make sure

that the alignment dot is in line with the upper pump tube track
as shown in

Figure 22

.

3. Hold the tubing with one hand while rotating the pump rotor to

guide the tubing around the rotor (

Figure 22

). The guide pins

will push the tubing into place.

4. Press the tubing into the lower pump tube track. Make sure that

the alignment dot is in line with the lower pump tube track as
shown in

Figure 22

.

5. Close and secure the pump cover with the thumbscrew.

6. Connect the open tube end to the tube fitting on the sampler

(

Figure 23

).

Figure 22 Pump tubing installation

1

Pump tube guide

5

Rotor

2

Pump tube track (upper)

6

Pump tube track (lower)

3

Guide pin

7

Connection to liquid sensor

4

Alignment dot