Sampler operation

4.9.3.4 Reset tubing cycles

When the “change tube” message is shown, replace the pump
tubing and then reset the pump cycles.

Procedure:

1. If the tubing life option is not enabled, select

SYSTEM SETUP

>

TUBING LIFE

>

ENABLE

>

RESET CYCLES

.

If the

tubing life option is enabled, select

SYSTEM SETUP

>

TUBING LIFE

>

RESET CYCLES

.

2. Select

YES

to reset the pump cycle count to 0. Press

ENTER

to

view the status screen to verify that the cycle count is set to 0.

Note: It is important to reset the pump cycles so that the change tube
message will be shown at the correct time for the new tube.