Calibration/verification routine – Luminex xPONENT for MAGPIX User Manual

Page 103

Advertising
background image

FIGURE 42.

Revive After Storate (Luminex) Routine

Calibration/Verification Routine

Calibration normalizes the settings for the system and ensures optimal and consistent
microsphere classification. Verification uses system controls to ensure that the analyzer is
functioning properly with current calibration settings.

1. Open the Home page.

2. On the Home page, click System Initialization under Daily Activities. The Auto Maint

tab opens.

3. Select the Calibration/Verification option.

4. Select the performance verification kit from the Performance Verification Kit list.

5. Select the calibration kit from the Calibration Kit list.

6. Vortex the xMAP calibrator and verification containers to ensure homogeneity. Do not

dilute xMAP calibrator and verification reagents.

7. Vortex the xMAP calibrator, verification, and fluidics containers to ensure homogeneity.

Do not dilute xMAP calibrator, verification, or fluidics agents.