Pause - Pauses the system after the current command completes. Pause does not stop the
system in the middle of running a command. You cannot run another command while the
system is paused.

Stop - Stops the system, regardless of command status.

Eject - Ejects the plate. Once the plate is ejected, the Eject button changes to Retract.
Retract retracts the plate, and the Retract button changes back to Eject.

Temp - Displays the difference in temperature between the current reading and the reading
when the system was calibrated, in degrees Celsius. If the temperature is out of tolerance,
this shows a high or low arrow. When clicked, it opens the Auto Maint tab.

XY Status - Displays the current location of the command, and the temperature of the plate
heating block in degrees Celsius. When clicked, it opens the Probe & Heater tab

Drive Fluid Level - The Drive Fluid liquid level sensor warns you when there is enough Drive
Fluid in the container to run one plate or less. The system stops if the container is empty.

Waste Fluid Level - The waste fluid container liquid level sensor warns you when the waste
fluid container is almost full. The system stops if the tank is full.

Home Page

The Home page displays a welcome message, batch creation buttons, Daily Activities
shortcuts, and the Installed Protocols list.