FIGURE 7.

Batch Options Tab

Batch Options. This section displays the following options:

Allow running a batch with expired reagents - Allows batches to run with expired

reagents.

Allow running a batch if XY temperature is not in range - Allows batches to run if the

XY temperature is not in range.

Allow running a batch if the instrument is not calibrated or verified - Allows batches to

run if the instrument is not calibrated.

Prompt for reacquiring of the last errored wells for an aborted partial batch - Allows

users to choose to reacquire aborted wells when resuming a stopped/partial batch or to
start with the next unacquired well.

Batch Settings This section displays the following options:

Calibration expiration days - Type the desired number of days a system calibration

remains active before expiring. The default number of days is seven.

Analysis display decimal places - Type the number of digits the system displays. The

default number of digits is three.

xPONENT for MAGPIX