3.2 Temperature

Calibration

For best DO accuracy, ensure that the
temperature is accurate. The temperature
sensor of the DO 6+ probe has been
factory calibrated, however if it changes
over time or if the probe is replaced,
calibration may be necessary. To protect
from erroneous calibrations, the allowable
tolerance is limited to ±5°C adjustment of
the factory default value.


Most users utilize automatic temperature
compensation using the sensor that is built
into the probe. However, manual
temperature compensation can be used to
input a fixed, known temperature.

1) Press

MODE to select temperature

mode. The display should show “T”
for temperature and “A” for automatic
temperature compensation.

2) Dip the probe into a solution with a

known, accurate temperature (i.e. a
temperature bath). Allow enough time
for the temperature reading to
stabilize.

3) Press

CAL. The CAL indicator will

blink above the display. The
temperature value shown is the value based on the factory default.

4) Press

▲ ▼ to adjust the reading to match the correct temperature value

(i.e. of the temperature bath).

5) Press

HOLD/ENTER to confirm the calibration and return to measurement

mode.

Press CAL as needed to exit calibration without confirmation at any time.

CAL

CAL

T

T

A

A

TA

TA

CAL

HOLD

ENTER

Factory default

Temperature
Measurement

“CO” will display for

1.5 seconds