Specifications – Nova-Tech DO 6+ User Manual

Page 40

Advertising
background image

Instruction Manual

DO 6+

- 34 -

10. SPECIFICATIONS

% Saturation Mode
Range
Resolution
Relative

accuracy


0.00 – 200.0 %
0.1 %
± 1.5% of Full Scale

mg/L (ppm) Concentration Mode
Range
Resolution
Relative

accuracy


0.00 – 20.00 mg/L or ppm
0.01 mg/L; 0.01 ppm
± 1.5% of Full Scale

Temperature
Range
Resolution
Relative

accuracy


-5.0 – 105.0 °C (meter only)*
0.1 °C
± 0.5 °C

*Probe measures 0.0 – 50.0 °C

Salinity Correction
Range
Resolution
Method


0.0 – 50.0 ppt
0.1 ppt
Automatic correction after manual input

Barometric Pressure Correction (mm Hg)
Range
Resolution
Method


500 to 1499 mm Hg or 66.6 to 199.9 kPA
1 mm Hg or 0.1 kPA
Automatic correction after manual input

Automatic Temperature Compensation

0.0 to 50.0°C

Manual Temperature Compensation

0.0 to 50.0°C

Probe (DO / Temp)

Galvanic / Thermistor

Probe Diameter

Body 12 mm, Cap 16 mm

Response Time

60 seconds to achieve 95% of the reading

% Saturation Calibration Points

100% in saturated air or air-saturated
water.

0% in zero oxygen solution

Advertising