Instruction Manual

DO 6+

- 23 -

5.5

Electrode Data (ELE.3)

Use this menu to view electrode data for diagnostic purposes. Data includes;
slope factor (FACT), % saturation offset (OFS), 100% saturation mV value
(HI.mV), and 0% saturation mV value (LO.mV). To view electrode data for %
saturation or mg/L (ppm) modes, enter from the corresponding setup menu.

The electrode slope factor gives an indication of the probe’s efficiency. It is the
ratio of the actual mV produced by the probe to the theoretical mV value. The
ratio displays from 0.5 to 1.999.

The % saturation offset allows you to view the electrode offset adjustment made
in Section 5.8 — % Saturation Offset Adjustment.

1) From the setup menu, press ▲▼ to select “ELE.3”.

2) Press

HOLD/ENTER to enter the menu. “FACT” is momentarily shown

before displaying the slope factor.

3) Press

HOLD/ENTER again. “OFS” is momentarily shown before displaying

% saturation offset. Note: this is not available from the mg/L (ppm)
concentration setup menu.

4) Press HOLD/ENTER again. “HI.mV” is momentarily shown before

displaying the electrodes mV output at 100%.

5) Press HOLD/ENTER again. “LO.mV” is momentarily shown before

displaying the electrodes mV output at 0%.

Press CAL to exit the setup mode and return to measurement mode at any
time.