Nova-Tech DO 6+ User Manual
Page 41

Instruction Manual
DO 6+
- 35 -
% Saturation Calibration Limits
Factory calibrated absolute value of 10.0%
and below for 0% point & 50% to 200% for
100% point.
Concentration Calibration Window
+/- 40% from the factory default
measurement value. Minimum reading
allowed is 2.00 mg/L (ppm).
Temperature Calibration Window
+/- 5°C from factory default measurement
Offset Adjustments (% Saturation)
+/- 10.0 of reading in Saturation mode
HOLD function
Yes
Auto-Off function
Selectable Auto Off function. (20 minutes
after last press)
Display
Custom Single 4 Digit LCD
Inputs
BNC for DO & 2.5 mm Phono for
temperature
Operating Range
0 to 50 °C
Power Requirements
4 AAA-sized batteries (included)
Battery Life
> 700 hours (Alkaline Batteries)
Dimensions
Meter: 15.7 x 8.5 x 4.2 cm / 255 g
Probe: 115 mm x 12 mm (Dia), 3-ft cable
Membrane housing: 16 mm (Dia)