Nova-Tech DO 6+ User Manual
Page 24

Instruction Manual
DO 6+
- 18 -
Setup Menu
Configuration Menu
A.ATC Automatic Temperature Compensation
b.DO Select mg/L or ppm units for concentration mode
(available from concentration setup only)
View the latest calibration data according to the respective setup
menu.
View the electrode properties
FACT View the slope factor
OFS View the % saturation offset adjustment.
(available from % saturation concentration setup only)
HI.mV View the mV value at 100% saturation
LO.mV View the mV value at 0% saturation
Selection of automatic power-off (20 min from the last key press)
Reset Menu
A.CAL Calibration only reset to factory default
b.USR User settings reset to factory default
% saturation offset adjustment
(available from % saturation concentration setup only)
Dissolved Oxygen Parameters
A.HG Barometric pressure adjustment in mmHg
A.PA Barometric pressure adjustment in kilopascal
b.SAL Salinity adjustment
(available from concentration setup only)