2 electrolyte solution, Troubleshooting – Nova-Tech DO 6+ User Manual

Page 37

Advertising
background image

Instruction Manual

DO 6+

- 31 -

6.2 Electrolyte

Solution

The electrolyte solution in your probe’s cap will deplete on usage and will need
to be replaced periodically. The replacement electrolyte solution included with
your probe comes premixed and ready to use. To order more electrolyte
solution, Section 11 – REPLACEMENTS AND ACCESSORIES.

7. TROUBLESHOOTING

Problem Cause

Solution

No display
when
turned on

a) Batteries not in place

b) Batteries not in correct

polarity (+ / – position)

c) Weak

batteries

a) Ensure batteries are in place and

making good contact

b) Re-insert batteries with correct

polarity

c) Replace batteries

Unstable
readings

a) Insufficient electrolyte in

probe

b) Air bubbles trapped around

the probe

c) Dirty or damaged probe

d) Probe not deep enough in

sample

e) External noise pickup or

induction caused by
nearby electric motor

f)

Broken probe

a)

Fill probe with electrolyte & / or
replace pre-assembled cap

b)

Stir or tap probe to remove
bubbles

c)

Clean the probe and re-
calibrate

d)

Make sure sample entirely
covers the probe sensors

e)

Move or switch off interfering
motor

f)

Replace probe

Slow
response

a) Dirty / Oily probe

b) Temperature

is

changing

a) Clean

probe

b)

Allow temperature to stabilize

No
response
to key
press

a)

HOLD in use- indicated
by flashing display

b)

Damaged pad

c)

Internal program error

a)

Press HOLD/ENTER

b) Return

to

dealer

c)

Reset by reinserting batteries