4 mg/l or ppm calibration – Nova-Tech DO 6+ User Manual

Page 19

Advertising
background image

Instruction Manual

DO 6+

- 13 -

3.4

mg/L or ppm Calibration

Performing a 100% saturation calibration will
simultaneously calibrate the corresponding
mg/L (ppm) concentration value. Therefore,
additional mg/L (ppm) calibration isn’t
required in most circumstances.

If desired, you can perform a calibration
adjustment in mg/L (ppm) without affecting
your % saturation calibration value.

1) Perform 100% saturation calibration.

See Section 3.3.1 — 100% Calibration.

2) Rinse the probe well with DI water or

rinse solution.

3) Dip the probe into a sample of known

oxygen concentration (i.e. determined
by titration or another instrument). Wait
for the reading to stabilize.

4) Press

MODE to select the mg/L (ppm).

5) Press

CAL. The CAL indicator and

current concentration display briefly,
before CAL flashes. The concentration
value based on the factory default
calibration is shown.

6) Press

▲ ▼ to adjust the reading to

match the known oxygen concentration
value.

7) Press HOLD/ENTER to confirm the

calibration. The meter displays “CO”, automatically calibrates to the
entered value, then returns to measurement mode.

NOTE: The minimum calibration value is 2 mg/L. To prevent erroneous
calibrations, the calibration is limited to ±40% adjustment of the factory default
value.

mg/l

mg/l

mg/l

CAL

CAL

HOLD

ENTER

CAL

mg/l

“CO” will display

for 1.5 seconds

A

A

A value with respect to the
factory default calibration
will be displayed