Nova-Tech DO 6+ User Manual

Page 18

Advertising
background image

Instruction Manual

DO 6+

- 12 -

3.3.2 0%

Calibration

1) Rinse the probe well with DI water or

rinse solution.

2) Press

MODE to select % saturation.

3) Dip the probe into zero oxygen

solution. Wait for the reading to
stabilize.

4) Press

CAL. The CAL indicator and

intended percentage calibration point
(0%) display briefly, before CAL
flashes. The % value based on the
factory default calibration is shown.

5) Press

HOLD/ENTER to confirm the

calibration. The meter displays “CO”,
automatically calibrates to 0.0%
saturation, then returns to
measurement mode.

A value with respect to the
factory default calibration

will be displayed

%

CAL

A

%

HOLD

ENTER

CAL

CAL

%

Calibration point will be
displayed for 1.5 seconds

%

“CO” will displayed for

1.5 seconds

A