5.8

% Saturation Offset Adjustment (OFS.6)

Use this feature to offset the meter’s value
when cross referenced to another DO
meter. The DO 6+ allows +/- 10.0% offset
adjustment. View the offset value from the
Electrode Data menu “ELE.3”.

1) Using the % saturation mode of your

DO6+, observe the reading a sample
solution after it has stabilized.

2) Similarly, observe the reading of the

same sample using another DO meter
as a reference. The probe of the
reference meter should be immersed
in the same sample at the same depth.

3) Switch off the DO6+ and enter the %

saturation setup menu.

4) Press

▲▼ to select “OFS.6”.

5) Press HOLD/ENTER to enter the

menu. The display will momentarily
show the measured reading based on
the last calibration before displaying
the last offset adjusted value.

6) Use

▲▼ to enter the value of the

reference DO meter.

7) Press

HOLD/ENTER to confirm.

8) Press

CAL to return to measurement mode.

Note: User calibrations will reset the offset adjustment to 0.0 %.

%

HOLD

ENTER

Measured reading based on last
percentage saturation calibration
is displayed momentarily

%

%

HOLD

ENTER

Last offset adjusted
value will be displayed

Display will return to “OFS.6” subgroup menu

“CO” will display for
1.5 seconds