Instruction Manual

DO 6+

- 14 -

%

A

Automatic Temperature
Compensation activated

4. MEASUREMENT

The DO 6+ offers automatic or manual temperature compensation.
See Section 5.2 — Automatic Temperature Compensation.

4.1 Temperature

Compensation

Insert the phono plug of the probe for Automatic Temperature Compensation
(ATC). The “A” annunciator will display normally when active, or blink if the
phono plug is disconnected.

Deactivate “A.ATC” for manual temperature
compensation. The “A” annunciator is not
displayed when the meter is in manual
temperature compensation mode.

For manual temperature compensation, manually
enter the desired process temperature into the
meter from 0 and 50 °C. Default is 25 °C.

1) Press

MODE to select temperature.

2) Press

CAL. The “CAL” indicator will start

blinking and the display will show the last
manually set temperature value.

3) Check the temperature of your sample using

an accurate reference thermometer. Wait for
the value to stabilize. Press ▲ ▼ to manually
set the temperature value.

4) Press HOLD/ENTER

to confirm the

calibration. The meter displays “CO”,
automatically sets the entered value, then
returns to measurement mode.

CAL

CAL

T

T

T

T

CAL

HOLD

ENTER

Last set manual
temperature setting

“CO” will display for 1.5

seconds