4.3

Pressure & Salinity Compensated Measurements

During measurement the dissolved oxygen reading is automatically
compensated for salinity and pressure based on values entered in the setup
menu. For best DO accuracy, ensure pressure and salinity are adjusted
accordingly from the setup menu. The factory default values are 760 mm Hg
(101.3 kPa) barometric pressure (sea level) and 0.0 ppt salinity (no salinity).
See Section 5.9 — Pressure & Salinity Adjustment.

4.4

Holding a Reading

To hold the displayed reading momentarily, press HOLD/ENTER during any
measurement. The %, mg/L, ppm or T annunciator will blink, indicating that the
value is held.

%

A

Annunciator Blinks when

meter is in “HOLD” mode

Press HOLD/ENTER again to deactivate the HOLD function. The meter reverts
to the current active measurement and the annunciator will stop blinking.

If the auto-off feature is activated, the meter will turn off automatically after 20
minutes of non use The HOLD value is not retained when the meter is shut off
automatically or manually.