Nova-Tech DO 6+ User Manual
Page 23
Advertising
Instruction Manual
DO 6+
- 17 -
5. SETUP
There are two setup menus. One menu is derived from the % saturation and
temperature modes, the other menu is derived from the mg/L (or ppm)
concentration mode.
1) Press MODE to display % or temperature. Press ON/OFF to power off.
OR
Press MODE to display mg/L (or ppm). Press ON/OFF to power off.
2) With the meter off, keep the MODE key pressed. Press and release
ON/OFF, then release MODE. The meter should display “SEt.P” after
ON/OFF is released and “COF.1” after MODE is released.
Press CAL one or more times as needed during setup mode to exit and return
to measurement mode at any time.
Advertising