5. SETUP

There are two setup menus. One menu is derived from the % saturation and
temperature
modes, the other menu is derived from the mg/L (or ppm)
concentration mode.

1) Press MODE to display % or temperature. Press ON/OFF to power off.


OR

Press MODE to display mg/L (or ppm). Press ON/OFF to power off.

2) With the meter off, keep the MODE key pressed. Press and release

ON/OFF, then release MODE. The meter should display “SEt.P” after
ON/OFF is released and “COF.1” after MODE is released.

Press CAL one or more times as needed during setup mode to exit and return
to measurement mode at any time.