3 % saturation calibration – Nova-Tech DO 6+ User Manual

Page 17

Advertising
background image

Instruction Manual

DO 6+

- 11 -

3.3

% Saturation Calibration

The DO6+ can be easily calibrated in air. For best DO accuracy, ensure that the
barometric pressure value is accurate.

The barometric pressure factory default is
760 mm Hg, which results in a theoretical
calibration value of 100% saturation in air.
If the barometric pressure setting has been
changed from 760 mm Hg, the meter will
automatically adjust to a new % saturation
calibration value instead of 100%. This new
value is correct for the adjusted barometric
pressure.

3.3.1

100% Calibration

1) Rinse the probe well with DI water or

rinse solution.

2) Press

MODE to select % saturation.

3) Hold the probe in the air with the tip

facing downwards. Wait for the
reading to stabilize.

4) Press

CAL. The CAL indicator and

intended percentage calibration point
(100%) display briefly, before CAL
flashes. The % value based on the
factory default calibration is shown.

5) Press

HOLD/ENTER to confirm the

calibration. The meter displays “CO”,
automatically calibrates to 100.0% air
saturation, then returns to
measurement mode.

%

CAL

A

%

HOLD

ENTER

CAL

CAL

%

Calibration point will be
displayed for 1.5 seconds

%

A value with respect to the
factory default calibration
will be displayed

“CO” will displayed for
1.5 seconds