The iDAS can be disabled, as opposed to suspended, only by disabling or deleting its individual data channels.

7.1.2. IDAS STRUCTURE

The iDAS is designed around the feature of a “record”. A record is a single data point. The type of data
recorded in a record is defined by two properties:

PARAMETER type that defines the kind of data to be stored (e.g. the average of gas concentrations

measured with three digits of precision). See Section 7.1.5.3.

A TRIGGER event that defines when the record is made (e.g. timer; every time a calibration is performed,

etc.). See Section 7.1.5.2.

The specific PARAMETERS and TRIGGER events that describe an individual record are defined in a construct
called a DATA CHANNEL (see Section 7.1.3). Each data channel is related one or more parameters with a
specific trigger event and various other operational characteristics related to the records being made (e.g. the
channels name, number or records to be made, time period between records, whether or not the record is
exported via the analyzer’s RS-232 port, etc.).

7.1.2.1. iDAS Channels

The key to the flexibility of the iDAS is its ability to store a large number of combinations of triggering events and
data parameters in the form of data channels. Users may create up to 50 data channels and each channel can
contain one or more parameters. For each channel, the following are selected:

 One triggering event is selected.
 Up to 50 data parameters, which can be the shared between channels.
 Several other properties that define the structure of the channel and allow the user to make operational

decisions regarding the channel.

Table 7-2: iDAS Data Channel Properties

PROPERTY DESCRIPTION

DEFAULT

SETTING

SETTING RANGE

NAME

The name of the data channel.

“NONE”

Up to 6 letters or digits

1.

TRIGGERING

EVENT

The event that triggers the data channel to measure
and store the datum.

ATIMER

Any available event

(see Appendix A-5).

NUMBER AND

LIST OF

PARAMETERS

A User-configurable list of data types to be
recorded in any given channel.

1

(COMEAS)

Any available parameter

(see Appendix A-5).

REPORT PERIOD

The amount of time between each channel data
point.

000:01:00

(1 hour)

000:00:01 to

366:23:59

(Days:Hours:Minutes)

NUMBER OF

RECORDS

The number of reports that will be stored in the data
file. Once the limit is exceeded, the oldest data is
over-written.

100

1 to 1 million, limited by

available storage space.

RS-232 REPORT

Enables the analyzer to automatically report
channel values to the RS-232 ports.

OFF

OFF or ON

CHANNEL

ENABLED

Enables or disables the channel. Allows a channel
to be temporarily turned off without deleting it.

ON

OFF or ON

CAL HOLD OFF

Disables sampling of data parameters while
instrument is in calibration mode

2

.

OFF

OFF or ON

1

More with APICOM, but only the first six are displayed on the front panel.

2

When enabled records are not recorded until the DAS HOLD OFF period is passed after calibration mode. DAS HOLD OFF SET in

the VARS menu (see Section

Error! Reference source not found.)

.