Analog output voltage / current range selection – Teledyne GFC-7001E - Trace CO Analyzer User Manual
Page 131

Advanced Features
Model GFC7001E Carbon Dioxide Analyzer
Teledyne Analytical Instruments
131
7.4.2. ANALOG OUTPUT VOLTAGE / CURRENT RANGE SELECTION
In its standard configuration, each of the analog outputs is set to output a 0–5 VDC signals. Several other output
ranges are available. Each range has is usable from -5% to + 5% of the rated span.
Table 7-7: Analog Output Voltage Range Min/Max
RANGE NAME
RANGE SPAN
MINIMUM OUTPUT
MAXIMUM OUTPUT
0.1V
0-100 mVDC
-5 mVDC
105 mVDC
1V
0-1 VDC
-0.05 VDC
1.05 VDC
5V
0-5 VDC
-0.25 VDC
5.25 VDC
10V
0-10 VDC
-0.5 VDC
10.5 VDC
The default offset for all VDC ranges is 0-5 VDC.
CURR
0-20 mA
0 mA
20 mA
While these are the physical limits of the current loop modules, typical applications use 2-20 or 4-20 mA for the lower and
upper limits. Please specify desired range when ordering this option.
The default offset for all current ranges is 0 mA.
Current outputs are available only on A1-A3.