Teledyne GFC-7001E - Trace CO Analyzer User Manual
Page 127
Advanced Features
Model GFC7001E Carbon Dioxide Analyzer
Teledyne Analytical Instruments
127
7.3. SETUP  MORE  DIAG: USING THE DIAGNOSTICS
FUNCTIONS
A series of diagnostic tools is grouped together under the SETUPMOREDIAG menu, as these parameters 
are dependent on firmware revision (see Appendix A). These tools can be used in a variety of troubleshooting 
and diagnostic procedures and are referred to in many places of the maintenance and trouble-shooting sections 
of this manual. 
The various operating modes available under the DIAG menu are:
Table 7-5: Diagnostic Mode (DIAG) Functions
DIAG SUBMENU
SUBMENU FUNCTION
Front Panel Mode
Indicator
MANUAL
SECTION
SIGNAL I/O
Allows observation of all digital and analog signals in 
the instrument. Allows certain digital signals such as 
valves and heaters to be toggled ON and OFF. 
DIAG I/O
13.1.3
ANALOG OUTPUT
When entered, the analyzer performs an analog 
output step test. This can be used to calibrate a 
chart recorder or to test the analog output accuracy. 
DIAG AOUT
13.5.7.1
ANALOG I/O
CONFIGURATION
This submenu allows the user to configure the 
analyzer’s analog output channels, including 
choosing what parameter will be output on each 
channel. Instructions that appear here allow 
adjustment and calibration of the voltage signals 
associated with each output as well as calibration of 
the analog to digital converter circuitry on the 
motherboard. 
DIAG AIO
7.4.1
ELECTRICAL
TEST
When activated, the analyzer performs an electrical 
test, which generates a voltage intended to simulate 
the measure and reference outputs of the 
SYNC/DEMOD board to verify the signal handling 
and conditioning of these signals. 
DIAG ELEC
9.6.4
13.5.6.2
DARK
CALIBRATION
1
Disconnects the preamp from synchronous 
demodulation circuitry on the SYNC/DEMOD PCA to 
establish the dark offset values for the measure and 
reference channel. 
DIAG DARK
CAL
9.6.1
PRESSURE
CALIBRATION
1
Allows the user to calibrate the sample pressure 
sensor. 
DIAG PCAL
9.6.2
FLOW
CALIBRATION
1
This function is used to calibrate the gas flow output 
signals of sample gas and ozone supply. 
DIAG FCAL
9.6.3
TEST CHAN
OUTPUT
Selects one of the available test channel signals to 
output over the A4 analog output channel. 
DIAG TCHN
7.4.6
1
These settings are retained after exiting DIAG mode.