3.3. ELECTRICAL CONNECTIONS

NOTE

To maintain compliance with EMC standards, it is required that the cable length be no greater than 3 meters for
all I/O connections, which include Analog In, Analog Out, Status Out, Control In, Ethernet/LAN, USB, RS-232,
and RS-485.

3.3.1. POWER CONNECTION

Attach the power cord to the analyzer and plug it into a power outlet capable of carrying at least 10 A current at
your AC voltage and that it is equipped with a functioning earth ground.

CAUTION

E

LECTRICAL

S

HOCK

H

AZARD

High Voltages are present inside the analyzer’s case.

Power connection must have functioning ground connection.

Do not defeat the ground wire on power plug.

Turn off analyzer power before disconnecting or

connecting electrical subassemblies.

Do not operate with cover off.

CAUTION

G

ENERAL

S

AFETY

H

AZARD

The GFC 7001E/EM Analyzer can be configured for both

100-130 V and 210-240 V at either 47 Hz or 63 Hz.

To avoid damage to your analyzer, make sure that the AC power voltage matches

the voltage indicated on the analyzer’s serial number label tag (See Figure 3-2)

before plugging the GFC 7001E/EM into line power.