Teledyne GFC-7001E - Trace CO Analyzer User Manual

Page 271

Advertising
background image

Troubleshooting & Repair

Model GFC7001E Carbon Dioxide Analyzer

Teledyne Analytical Instruments

271

The following table contains some of the more common causes for these values to be out of range.

Table 13-2: Test Functions - Indicated Failures

TEST

FUNCTIONS

(

As Displayed)

INDICATED FAILURE(S)

TIME

Time of day clock is too fast or slow.
To adjust, see Section 6.5.4.
Battery in clock chip on CPU board may be dead.

RANGE

Incorrectly configured measurement range(s) could cause response problems with a Data logger or chart
recorder attached to one of the analog output.
If the Range selected is too small, the recording device will over range.
If the Range is too big, the device will show minimal or no apparent change in readings.

STABIL

Indicates noise level of instrument or CO concentration of sample gas (see Section 13.4.2 for causes).

CO MEAS

&

CO REF

If the value displayed is too high the IR Source has become brighter. Adjust the variable gain potentiometer on
the sync/demod board (see Section 13.5.6.1).
If the value displayed is too low or constantly changing and the CO REF is OK:

 Failed multiplexer on the mother board
 Failed sync/demod board
 Loose connector or wiring on sync/demod board
 If the value displayed is too low or constantly changing and the CO REF is bad:
 GFC Wheel stopped or rotation is too slow
 Failed sync/demod board IR source
 Failed IR source
 Failed relay board
 Failed I

2

C bus

 Failed IR photo-detector

MR Ratio

When the analyzer is sampling zero air and the ratio is too low:

 The reference cell of the GFC Wheel is contaminated or leaking.
 The alignment between the GFC Wheel and the segment sensor, the M/R sensor or both is incorrect.
 Failed sync/demod board

When the analyzer is sampling zero air and the ratio is too high:

 Zero air is contaminated
 Failed IR photo-detector

PRES

See Table 13-1 for SAMPLE PRES WARN.

SAMPLE FL

Check for gas flow problems (see Section 13.2).

SAMP TEMP

SAMPLE TEMP should be close to BENCH TEMP. Temperatures outside of the specified range or oscillating
temperatures are cause for concern.

BENCH

TEMP

Bench temp control improves instrument noise, stability and drift. Temperatures outside of the specified range
or oscillating temperatures are cause for concern. Table 13-1 for BENCH TEMP WARNING.

WHEEL

TEMP

Wheel temp control improves instrument noise, stability and drift. Outside of set point or oscillating
temperatures are causes for concern. See Table 13-1 for WHEEL TEMP WARNING.

BOX TEMP

If the box temperature is out of range, check fan in the power supply module. Areas to the side and rear of
instrument should allow adequate ventilation. See Table 13-1 for BOX TEMP WARNING.

Advertising