Teledyne GFC-7001E - Trace CO Analyzer User Manual

Page 134

Advertising
background image

Advanced Features

Model GFC7001E Carbon Dioxide Analyzer

Teledyne Analytical Instruments

134

7.4.3.2. Automatic Calibration of the Analog Outputs

To calibrate the outputs as a group with the AOUTS CALIBRATION command, select the ANALOG I/O
CONFIGURATION
submenu (see Figure 7-4) then press:

NOTE

Before performing this procedure, make sure that the AUTO CAL for each analog output is enabled.

(See Section 7.4.3.1)