Teledyne GFC-7001E - Trace CO Analyzer User Manual

Page 52

Advertising
background image

Getting Started

Model GFC7001E Carbon Dioxide Analyzer

Teledyne Analytical Instruments

52

3.6.2.1. Verifying the GFC 7001E/EM Reporting Range Settings

While it is possible to perform the following procedure with any range setting we recommend that you perform
this initial checkout using following reporting range settings:

 Unit of Measure: PPM
 Analog Output Reporting Range: 50 ppm
 Mode Setting: SNGL

While these are the default setting for the GFC 7001E/EM Analyzer, it is recommended that you verify them
before proceeding with the calibration procedure, by pressing:

SETUP X.X

RANGE MODE:SINGL

SNGL DUAL AUTO

ENTR

EXIT

SAMPLE RANGE=50.0 PPM CO= XX.XX

<TST

TST> CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK MORE

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET UNIT

DIL

EXIT

SETUP X.X

RANGE: 50.0 Conc

0

0

0

5

0

.0

ENTR

EXIT

SETUP X.X

CONC UNITS:PPM

PPB PPM UGM MGM

ENTR

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET

UNIT

DIL

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET UNIT

DIL

EXIT

Verify that the MODE

is set for SNGL.

If it is not, press

SINGL ENTR.

Press EXIT

3x’s to return

the M200E to

the SAMPLE

mode.

Verify that the RANGE is

set for 50.0

If it is not, toggle each

numeric key until the

proper range is set, then

press ENTR.

Verify that the UNITs

is set for PPM

If it is not, press

PPM ENTR.