Getting Started

Model GFC7001E Carbon Dioxide Analyzer

Teledyne Analytical Instruments

49

Table 3-9: Possible Warning Messages at Start-Up

Message MEANING

ANALOG CAL WARNING

The instrument's A/D circuitry or one of its analog outputs is not calibrated.

BENCH TEMP WARNING

Optical bench temperature is outside the specified limits.

BOX TEMP WARNING

The temperature inside the GFC 7001E/EM chassis is outside the specified
limits.

CANNOT DYN SPAN

2

Remote span calibration failed while the dynamic span feature was set to
turned on.

CANNOT DYN ZERO

3

Remote zero calibration failed while the dynamic zero feature was set to turned
on.

CONFIG INITIALIZED

Configuration was reset to factory defaults or was erased.

DATA INITIALIZED

iDAS data storage was erased.

FRONT PANEL WARN

CPU is unable to communicate with the front panel.

PHOTO TEMP WARNING

Photometer temperature outside of warning limits specified by
PHOTO_TEMP_SET variable.

REAR BOARD NOT DET

Motherboard was not detected during power up.

RELAY BOARD WARN

CPU is unable to communicate with the relay PCA.

SAMPLE FLOW WARN

The flow rate of the sample gas is outside the specified limits.

SAMPLE PRESS WARN

Sample pressure outside of operational parameters.

SAMPLE TEMP WARN

The temperature of the sample gas is outside the specified limits.

SOURCE WARNING

The IR source may be faulty.

SYSTEM RESET

1

The computer was rebooted.

WHEEL TEMP WARNING

The Gas Filter Correlation Wheel temperature is outside the specified limits.

1

Clears 45 minutes after power up.

2

Clears the next time successful zero calibration is performed.

3

Clears the next time successful span calibration is performed.