The status outputs are accessed via a 12-pin connector on the analyzer’s rear panel labeled STATUS (see
Figure 3-2). Pin-outs for this connector are:

STATUS

1 2 3 4 5 6 7 8 D

+

SYST

EM OK

HIGH RANG

E

C

O

NC V

A

L

ID

Z

E

RO

CAL

SP

A

N

CAL

DIAG

M

O

D

E

Opt

io

n

al O

2

CAL

Figure 3-8:

Status Output Connector

Table 3-5: Status Output Signals

REAR PANEL

LABEL

STATUS

DEFINITION

CONDITION

1

SYSTEM OK

ON if no faults are present.

2 CONC

VALID

OFF any time the HOLD OFF feature is active, such as during calibration or when
other faults exist possibly invalidating the current concentration measurement
(example: sample flow rate is outside of acceptable limits).
ON if concentration measurement is valid.

3 HIGH

RANGE

ON if unit is in high range of either the DUAL or AUTO range modes.

4 ZERO

CAL

ON whenever the instrument’s ZERO point is being calibrated.

5 SPAN

CAL

ON whenever the instrument’s SPAN point is being calibrated.

6 DIAG

MODE

ON whenever the instrument is in DIAGNOSTIC mode.

7 CO

2

CAL

If this analyzer is equipped with an optional CO

2

sensor, this Output is ON when that

sensor is in calibration mode.
Otherwise this output is unused.

8 O

2

CAL

If this analyzer is equipped with an optional O

2

sensor, this Output is ON when that

sensor is in calibration mode.
Otherwise this output is unused.

D

EMITTER BUS

The emitters of the transistors on pins 1-8 are bussed together.

SPARE

+

DC POWER

+ 5 VDC, 300 mA source (combined rating with Control Output, if used).

Digital Ground

The ground level from the analyzer’s internal DC power supplies.