Teledyne GFC-7001E - Trace CO Analyzer User Manual
Page 40

Getting Started
Model GFC7001E Carbon Dioxide Analyzer
Teledyne Analytical Instruments
40
The status outputs are accessed via a 12-pin connector on the analyzer’s rear panel labeled STATUS (see
Figure 3-2). Pin-outs for this connector are:
STATUS
1 2 3 4 5 6 7 8 D
+
SYST
EM OK
HIGH RANG
E
C
O
NC V
A
L
ID
Z
E
RO
CAL
SP
A
N
CAL
DIAG
M
O
D
E
Opt
io
n
al O
2
CAL
Figure 3-8:
Status Output Connector
Table 3-5: Status Output Signals
REAR PANEL
LABEL
STATUS
DEFINITION
CONDITION
1
SYSTEM OK
ON if no faults are present.
2 CONC
VALID
OFF any time the HOLD OFF feature is active, such as during calibration or when
other faults exist possibly invalidating the current concentration measurement
(example: sample flow rate is outside of acceptable limits).
ON if concentration measurement is valid.
3 HIGH
RANGE
ON if unit is in high range of either the DUAL or AUTO range modes.
4 ZERO
CAL
ON whenever the instrument’s ZERO point is being calibrated.
5 SPAN
CAL
ON whenever the instrument’s SPAN point is being calibrated.
6 DIAG
MODE
ON whenever the instrument is in DIAGNOSTIC mode.
7 CO
2
CAL
If this analyzer is equipped with an optional CO
2
sensor, this Output is ON when that
sensor is in calibration mode.
Otherwise this output is unused.
8 O
2
CAL
If this analyzer is equipped with an optional O
2
sensor, this Output is ON when that
sensor is in calibration mode.
Otherwise this output is unused.
D
EMITTER BUS
The emitters of the transistors on pins 1-8 are bussed together.
SPARE
+
DC POWER
+ 5 VDC, 300 mA source (combined rating with Control Output, if used).
Digital Ground
The ground level from the analyzer’s internal DC power supplies.