Instrument id code – Teledyne GFC-7001E - Trace CO Analyzer User Manual
Page 183

Remote Operation
Model GFC7001E Carbon Dioxide Analyzer
Teledyne Analytical Instruments
183
To assign or reset the status flag bit assignments, press:
<TST
TST> CAL
deletes the
character currently
inside the cursor
brackets.
Press the
key repeatedly to cycle
through the available character set:
: Values of
can also be set
but are meaningless.
The
and
keys move the cursor
brackets “
left and right along the
bit string.
SETUP X.X
ID
COM1 COM2
EXIT
SETUP X.X
CFG DAS RNGE PASS CLK
EXIT
SETUP X.X
VARS DIAG
EXIT
Continue pressing
until ...
SETUP X.X
EDIT PRNT EXIT
SETUP X.X
SETUP X.X
PREV NEXT
PRNT EXIT
Continue pressing
until desired
flag message is displayed
discards the
new setting.
accepts the
new setting.
8.6.8. INSTRUMENT ID CODE
Each instrument on a Hessen Protocol network must have a unique ID code. If more than one GFC 7001E/EM
Analyzer is on the Hessen network, you will have to change this code for all but one of the GFC 7001E/EM
Analyzer’s on the Hessen network (see Section 8.1.6).
The default ID code for the GFC 7001E/EM Analyzers is 300.