Epa equivalency designation – Teledyne GFC-7001E - Trace CO Analyzer User Manual

Page 28

Advertising
background image

Specifications

Model GFC7001E Carbon Dioxide Analyzer

Teledyne Analytical Instruments

28

1

As defined by the USEPA

2

At constant temperature and pressure

2.2. EPA EQUIVALENCY DESIGNATION

Teledyne’s GFC 7001E Carbon Monoxide Analyzer is designated as Reference Method Number EQOA-0992-
087 as defined in 40 CFR Part 53, when operated under the following conditions:

 Range: Any range from 10 ppm to 50 ppm.
 Ambient temperature range of 10 to 40C.
 Line voltage range of 90 – 127 and 200 – 230 VAC, 50/60 Hz.
 Sample filter: Equipped with PTFE filter element in the internal filter assembly.
 Sample flow of 800  80 cm

3

/min at sea level.

 Internal sample pump.
 Software settings:

Dilution factor

1.0

AutoCal

ON or OFF

Dynamic Zero

ON or OFF

Dynamic Span

OFF

Dual range

ON or OFF

Auto range

ON or OFF

Temp/Pres compensation

ON

Under the designation, the analyzer may be operated with or without the following options:

 Rack mount with slides.
 Rack mount without slides, ears only.
 Zero/span valve options.

 Option 50A – Sample/Cal valves, or;
 Option 50B – Sample/Cal valves with span shutoff & flow control.

 Internal zero/span (IZS) option with either:

 Option 51A – Sample/Cal valves, or;
 Option 51C – Sample/Cal valves with span shutoff & flow control.

 Status outputs.
 Control inputs.
 RS-232 output.
 Ethernet output.
 4-20mA, isolated output.