Tuv designation, Ce mark compliance, Emissions compliance – Teledyne GFC-7001E - Trace CO Analyzer User Manual

Page 29: Safety compliance

Advertising
background image

Specifications

Model GFC7001E Carbon Dioxide Analyzer

Teledyne Analytical Instruments

29

2.3. TUV DESIGNATION

On behalf of Teledyne TÜV Rheinland Immissionsschutz und Energiesysteme GmbH has performed the
suitability test of the measuring system GFC 7001E for the component carbon monoxide.

The suitability test was carried out in compliance with the following guidelines and requirements:

 EN 14626 Ambient Air Quality – Standard method for the measurement of the concentration of carbon

monoxide by nondispersive infrared spectroscopy, March 2005.

The measuring system GFC 7001E operates using the non-dispersive infrared spectroscopy.

The investigations have been carried out in the laboratory and during a field test, lasting three months. The
tested measuring ranges are:

Component Measuring

Range

Carbon Monoxide

CO

100

mg/m

3

EN

14626

NOTE: 0-100 ppm correlates to 0-100 µmol/mol or 0-116 mg/m

3

(at 293 K and 1013

mbar).

The minimum requirements have been fulfilled in the suitability test.

Therefore the TÜV Immissionsschutz and Energiesysteme GmbH proposes the publication as a suitability-tested
measuring system for continuous monitoring of carbon monoxide in the ambient air.

2.4. CE MARK COMPLIANCE

2.4.1. EMISSIONS COMPLIANCE

Teledyne’s GFC 7001E/EM Gas Filter Correlation CO Analyzer was tested and found to be fully compliant with:

EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B section 15.107 Class A, ICES-003 Class A (ANSI
C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97) Class A.

Tested on 11-29-2001 at CKC Laboratories, Inc., Report Number CE01-249.

2.4.2. SAFETY COMPLIANCE

Teledyne’s GFC 7001E/EM Gas Filter Correlation CO Analyzer was tested and found to be fully compliant with:

IEC 61010-1:90 + A1:92 + A2:95,

Tested on 02-06-2002 at NEMKO, Report Number 2002-012219.