Pressure calibration – Teledyne GFC-7001E - Trace CO Analyzer User Manual

Page 206

Advertising
background image

Calibration Procedures

Model GFC7001E Carbon Dioxide Analyzer

Teledyne Analytical Instruments

206

9.6.2. PRESSURE CALIBRATION

A sensor at the exit of the sample chamber continuously measures the pressure of the sample gas. This data is
used to compensate the final CO concentration calculation for changes in atmospheric pressure and is stored in
the CPU’s memory as the test function PRES (also viewable via the front panel).

NOTE

This calibration must be performed when the pressure of the sample gas is equal to ambient

atmospheric pressure.

Before performing the following pressure calibration procedure, disconnect the sample gas pump and

the sample gas-line vent from the sample gas inlet on the instrument’s rear panel.

To cause the analyzer to measure and record a value for PRES, press.

SETUP X.X

CFG DAS RNGE PASS CLK

EXIT

SETUP X.X

COMM VARS

EXIT

SETUP X.X

EXIT

ENTR

Continue pressing

until ...

discards the

new setting.

accepts the

new setting.

DIAG

PREV NEXT

EXIT

DIAG PCAL

EXIT

DIAG PCAL