Pressure calibration – Teledyne GFC-7001E - Trace CO Analyzer User Manual
Page 206

Calibration Procedures
Model GFC7001E Carbon Dioxide Analyzer
Teledyne Analytical Instruments
206
9.6.2. PRESSURE CALIBRATION
A sensor at the exit of the sample chamber continuously measures the pressure of the sample gas. This data is
used to compensate the final CO concentration calculation for changes in atmospheric pressure and is stored in
the CPU’s memory as the test function PRES (also viewable via the front panel).
NOTE
This calibration must be performed when the pressure of the sample gas is equal to ambient
atmospheric pressure.
Before performing the following pressure calibration procedure, disconnect the sample gas pump and
the sample gas-line vent from the sample gas inlet on the instrument’s rear panel.
To cause the analyzer to measure and record a value for PRES, press.
SETUP X.X
CFG DAS RNGE PASS CLK
EXIT
SETUP X.X
COMM VARS
EXIT
SETUP X.X
EXIT
ENTR
Continue pressing
until ...
discards the
new setting.
accepts the
new setting.
DIAG
PREV NEXT
EXIT
DIAG PCAL
EXIT
DIAG PCAL