AORM Software Package

923133 Rev A

ISSUED:

June 2013

21

4. Touch

inside

the

Operator1

field and select Trend

from the Select Math

Operator

pop-up menu. The "Trend" setup dialog will appear at the right of the screen:

5. Decide whether all the parameter values, all per trace, or only the average of all

parameter calculations for each waveform acquisition should be placed in the trend.

All

-- every parameter calculation on each waveform will be placed in the trend.

Average

-- trends only the average of all the values calculated on a given acquisition and

yields one point in the trend per acquisition.
All per Trace

-- for each acquisition, clears the buffer and places all parameter

calculations from the new data in the trend. Unless this is specifically required, All should
be selected.

6. Choose the number of values to be placed in the generated trend.

7. If desired, you can also configure the center and height of the trend in the base units of

the parameter being trended. However, this is not a requirement, and Find Scale can be
used to center the trend after it has been calculated.

Center

is for selecting the mantissa, exponent, or number of digits resolution, using the

associated knob. The configuration is the value at the horizontal center line on the grid,
while units are those of the parameter trended.

Height/div

selects the vertical value of each vertical screen division. Units are those of

the parameter trended.