Pmoda, Pit modulation amplitude, Description – Teledyne LeCroy AORM - Advanced Optical Recording Measurements User Manual

Page 70: Display options, Pmoda pit modulation amplitude, Pmoda paa avg top

Advertising
background image

68

ISSUED:

June 2013

923133 Rev A

PMODA PIT

MODULATION

AMPLITUDE

Description

Pit Modulation Amplitude provides a measurement of the ratio of the Pit Average Amplitude (paa)
for the low ‘nT’ pits/spaces in the data signal to the Pit Top (ptop) of the high ‘nT’ pits in the data
signal:

PMODA

paa

avg top

low n

high n

=

_

_

(

)

The low and high ‘nT’ values to be used for performing the calculation are provided by the user
through the associated measurement configuration options. Some measurements of modulation
amplitude require the low and high n index to be identical. The value is shown is decimal. The
measurement calculation is compliant with the definition of Modulation Amplitude as defined by
IEC 908:1987 Section 9.2 and ISO/IEC 10149:1995 (E) Section 12.2.

Note: This measurement must be performed on the DC-coupled optical data waveform, otherwise incorrect values will
result.

Display Options

ORM parameter calculations can be displayed, histogrammed, and trended in a variety of ways.

DISPLAY TYPE

VALUE DISPLAYED

Parameter Statistics Off

Single value of the modulation amplitude for the last acquisition.

Parameter Statistics On

Average, minimum, maximum, and sigma of the single
modulation amplitude value calculated per acquisition for all
acquisitions since the last C

LEAR

S

WEEPS

operation.

Histogram Function

Histogram of the modulation amplitude value calculated per
acquisition for all acquisitions since the last C

LEAR

S

WEEPS

operation.

Trend Function

Trend of the single modulation amplitude value calculated per
acquisition for all acquisitions since the last C

LEAR

S

WEEPS

operation.