Ptop, Pit top, Description – Teledyne LeCroy AORM - Advanced Optical Recording Measurements User Manual

Page 79: Display options, Aorm software package, Ptop pit top

Advertising
background image

AORM Software Package

923133 Rev A

ISSUED:

June 2013

77

PTOP PIT

TOP

Description

Pit Top provides a measurement of the best estimate of the top amplitude of a pit. The concept of the top
calculation is to automatically provide the same measurement which would be obtained from a persistence
plot. The top of each pit is determined through histogramming techniques described in detail under Base and
Top Calculation Details.

When ptop is configured as a custom parameter, all tops within the single nT or

range of nT are calculated. Histogramming or trending such a configuration would result in one value per pit in
the nT range contributing a value to the histogram or trend. The value displayed on the custom parameter
display line is the average of all such top calculations. ‘nT Table’ mode provides an average top measurement
for each n index.

Display Options

ORM parameter calculations can be displayed, histogrammed, and trended in a variety of ways.

DISPLAY TYPE

VALUE DISPLAYED

Parameter Statistics Off

All values of the top for all pits within the selected ‘nT’ range for the last
acquisition.

Parameter Statistics On

Average, minimum, maximum, and sigma of the top for all pits within
the selected ‘nT’ range for all acquisitions since last C

LEAR

S

WEEPS

operation.

nT Table

List of values of the average top for each group of pits of common ‘nT’
width for the last acquisition.

Histogram Function

Histogram graph of the value of the top for all pits within the selected
‘nT’ range for all acquisitions since the last C

LEAR

S

WEEPS

operation.

Trend Function

Trend graph of the value of the top calculated for pit that is within the
selected ‘nT’ range for all acquisitions since last C

LEAR

S

WEEPS

operation.