Using parameters with trends and xy plots, Example and step-by-step instructions – Teledyne LeCroy AORM - Advanced Optical Recording Measurements User Manual

Page 98

Advertising
background image

96

ISSUED:

June 2013

923133 Rev A

USING PARAMETERS WITH TRENDS AND XY PLOTS

X-axis

Y-axis

t@pit Dp2c

edgsh
pbase
pmax
pmin
ptop
pwid

pwid pbase

pmax
pmin
ptop

ptop pmax

pbase pmin

We saw in the t@pit parameter description how the ORM parameters have
certain unique characteristics that make particular measurements useful
when trended together with XY plots. And how the t@pit parameter is
essential to those measurements. Plots that can be generated on single
acquisitions include those listed in the table at left.
The reason that these plots are considered useful on single acquisitions is
because the parameters are guaranteed to be configurable in a manner
that meets the following criteria:

Each parameter is capable of providing multiple values per acquisition.

Each parameter pair is configurable in a manner that guarantees the
same number of events per waveform.

Each parameter pair must be configurable in a manner that
guarantees a one-to-one correspondence between parameter
calculation values.

Example and Step-by-Step Instructions

Here is an example typifying the use of XY plots without the t@pit parameter. A complete
example using t@pit has been provided in the section dedicated to this parameter description.

Consider a situation in which it is desirable to find the relationship of the pit top value to the pit
width in an optical recognition data waveform:

The scope is set up to acquire this waveform by assigning Channel 1 to the data signal at a
time/div of 0.2 ms. This signal will contain approximately 1800 pits/spaces. The ideal threshold
has been determined by the pmidl parameter as 1.9 mV.

The ptop custom parameter is assigned to P1, and the pwid parameter is assigned to P2. Use
configuration tracking to configure both parameters in the following manner:

hysteresis = 0.5 divisions
threshold = 1.9 mV
polarity = Pos
range of n
low n = 3
high n = 11
period = 231.5 ns

In math setup, F1 is defined as the trend of the ptop parameter and F2 as the trend of the pwid
parameter. Because we are expecting 1800 pits/spaces, make sure that for each math setup the
trends are set to use up to 20,000 values, the maximum amount.

Note:

If configuration tracking is used on the ptop parameter, the pwid parameter must be visited in order to set the

polarity to positive because ptop inherently implies positive polarity pits.

Advertising