Pasym, Pit asymmetry, Description – Teledyne LeCroy AORM - Advanced Optical Recording Measurements User Manual

Page 60: Display options, Pasym pit asymmetry, Pasym pmidl pmidl paa

Advertising
background image

58

ISSUED:

June 2013

923133 Rev A

PASYM PIT

ASYMMETRY

Description

Pit Asymmetry provides a measurement of the asymmetry of the middle voltage level for the high
nT index pits/spaces compared to the middle voltage level of the low ‘nT’ index pits/spaces. The
measurement calculation is compliant with the definition of Pit Asymmetry as defined by IEC
908:1987 Section 3.1. The negative value of the measurement is referred to as Pit Symmetry as
defined by ISO/IEC 10149:1995 (E) Section 12.2. Pit Asymmetry is calculated by the formula:

PASYM

pmidl

pmidl

paa

high n

low n

high n

=

_

_

_

100%

where paa is the average peak-peak amplitude. The low (smallest) and high (largest) ‘nT’ values to
use in performing the calculation are provided by the user through the associated measurement
configuration options. Midpoint designates the midpoint value between the average top and base for a
specified ‘nT.’ The value shown is in units of percent.

Display Options

ORM parameter calculations can be displayed, histogrammed, and trended in a variety of ways.

DISPLAY TYPE

VALUE DISPLAYED

Parameter Statistics Off

Single value of the asymmetry for the last acquisition.

Parameter Statistics On

Average, minimum, maximum and sigma of the single
asymmetry value calculated per acquisition for all acquisitions
since the last C

LEAR

S

WEEPS

operation.

Histogram Function

Histogram of asymmetry value calculated per acquisition for all
acquisitions since last C

LEAR

S

WEEPS

operation.

Trend Function

Trend of single asymmetry value calculated per acquisition for
all acquisitions since last C

LEAR

S

WEEPS

operation.