AORM Software Package

923133 Rev A

ISSUED:

June 2013

31

Pit or Space Identification

This is determined uniquely by the threshold, hysteresis, and edge polarity of threshold crossings.
A positive threshold crossing indicates the start of a positive polarity pit and the end of a negative
polarity space. A positive threshold crossing followed by a negative threshold crossing fully
delineates a pit. A negative crossing followed by a positive crossing fully delineates a space, as
illustrated in the following figure.

In order to prevent false pit and space identifications, hysteresis is provided. Hysteresis adds an
additional condition that must be met before a threshold crossing is recognized as a pit/space
edge. It requires that the waveform make an excursion of a certain distance from the threshold
before the next threshold crossing is recognized.

The next figure shows a threshold crossing that would result in incorrect pit identification without
hysteresis.

Voltage

Threshold

Space

Pit

False

Pit

Level

Threshold