Example, Aorm software package – Teledyne LeCroy AORM - Advanced Optical Recording Measurements User Manual

Page 55

Advertising
background image

AORM Software Package

923133 Rev A

ISSUED:

June 2013

53

DISPLAY TYPE

VALUE DISPLAYED

Parameter Statistics Off

All values of the overall edge shift for all pits/spaces within the
selected ‘nT’ range for last acquisition.

Parameter Statistics On

Average, minimum, maximum, and sigma of all edge shift
values calculated per acquisition for all acquisitions since the
last C

LEAR

S

WEEPS

operation.

nT Table

List of values of the overall edge shift for each group of
pits/spaces of common ‘nT’ width for the last acquisition.

Histogram Function

Histogram of the single overall edge shift value calculated for
each acquisition for all acquisitions since the last C

LEAR

S

WEEPS

operation.

Trend Function

Trend of the single overall Edge Shift value calculated for each
acquisition for all acquisitions since the last C

LEAR

S

WEEPS

operation.

Example

The example shows the CD data signal measured at the selected voltage threshold containing, in
sequence, a 5T pit, 3T space, 3T pit and 4T space. If the clock period ‘T’ is 231.5 ns, then the 5T
and 4T edge shift value is simply the difference between the width calculated and the ideal width
(since there is only one pit/space of that ‘nT’ width), thus:

(

)

(

)

edgsh T

edgsh T

(

)

.

.

.

(

)

.

.

.

4

920

4 2315

100%

2315

2 59%

5

1160 5 2315

100%

2315

1 08%

=

− ⋅

= −

=

− ⋅

= +

The 3T edge shift value is the average difference:

(

)

(

)

edgsh T

(

)

.

.

.

.

.

3

690

3 2315

100%

2315

695 3 2315

100%

2315

2

0 86%

=

− ⋅

⎝⎜

⎠⎟ +

− ⋅

⎝⎜

⎠⎟

= +

In an nT Table display, these three values would be shown in the appropriate nT location.