f. The following screen will be displayed:

Press and hold down the

[ARM]

switch, and then press the

[START]

key. Continue to

hold down the

[ARM]

switch.

g. The following screen will be displayed during testing:

Continue to hold down the

[ARM]

switch until the test is completed.

h. Once the test is completed, you will be returned to the “START-UP” menu.

Please see section 3.7 for instructions on how to print the test results.

NOTE

The DigiTMR can be configured to print tabulated and graphic reports
automatically after a test is completed. Please see section 3.10.4 for further
information.

OPEN TIMING
Hold “ARM” switch,
then “START”.
“STOP” to ABORT

Test in Progress
Hold “ARM” until
test complete.
(Up to 25 seconds)