NIBP System Messages

When monitoring NIBP, the Propaq MD unit may display the following messages:

System Message

Cause

READING IN PROGRESS

The unit is taking an NIBP measurement and
functioning normally.

READING STOPPED

The unit has stopped an NIBP measurement,
because the operator has pressed the NIBP button
and cancelled the measurement.

READING FAILED

The patient’s pulse is too weak to for NIBP
measurement, or the cuff requires adjustment.

ARTIFACT

Excessive artifact is preventing the NIBP
measurement.

NEONATE ALERT

The unit has detected a neonate cuff when in Adult
mode. Replace cuff or correct patient mode, as
appropriate.

AIR LEAK

A major air leak is preventing cuff inflation. Check
hose and cuff connections, replace a defective hose
or cuff, as necessary, and reattempt NIBP
measurement.

KINKED HOSE

An air obstruction is preventing the correct operation
of the unit. Check hose for kinks. Eliminate kinks or
replace hose, if necessary.

NIBP DISABLED

A system error has occurred, and the Propaq MD
unit should be serviced.

NIBP CALIBRATION CHECK DUE The annual calibration check is due. You can

continue to use the device, however, we recommend
that you contact the ZOLL Technical Service
Department or a ZOLL authorized service
representative as soon as possible to perform the
recommended annual NIBP calibration check.