Connecting the spo, Sensor – ZOLL Propaq MD Rev A User Manual
Page 118
Advertising
C
HAPTER
9 M
ONITORING
P
ULSE
O
XIMETRY
(S
P
O2)
9-6
www.zoll.com
9650-0805-01 Rev. A
Connecting the SpO
2
Sensor
To connect the SpO
2
sensor to the Propaq MD unit:
1. When using a sensor extension cable, inspect the cable before use. Replace the cable if it
shows any signs of wear, breakage, or fraying. Plug the sensor extension cable into the SpO
2
socket on the side of the Propaq MD unit:
Figure 9-1 Connecting the SpO
2
Sensor to the Propaq MD Unit
2.
Lift the clear plastic protective cover from the female end of the extension cable, then plug
the sensor cable’s male connector into the extension cable connector -- make sure that the
plug is all the way into the connector:
Figure 9-2 Plug the Sensor Connectors Together