When you are finished changing values on the alarm menu, navigate to the Backarrow key
to exit the menu.

Setting Upper and Lower Systolic Alarm Limits

Initially, the SpO

2

Alarm Settings menu specifies whether the SpO

2

alarms are enabled (

ON

)

or

disabled (

OFF

),

and displays the default upper and lower systolic limits. The upper and lower

limits can be

ON

or

OFF

(default is

OFF

).The following table lists the default SpO

2

limits for

adult, pediatric, and neonate patients, and gives the range in which you can set these limits.

Using the SpO

2

Parameter Control Panel

Use the navigation keys to highlight and select the SpO

2

numeric display to display the

SpO

2

Parameter Control Panel, on which you can select options that optimize the use of SpO

2

for the patient:

Figure 9-4 SpO

2

Parameter Control Menu

Patient Type

SpO

2

Limit Default

SpO

2

Limit Range

Adult

Lower: 85%

Upper: 100%

Lower: 50 - 98%

Upper: 52 - 100%

Pediatric

Lower: 85%

Upper: 100%

Lower: 50 - 98%

Upper: 52 - 100%

Neonate

Lower: 85%

Upper: 95%

Lower: 50 - 98%

Upper: 52 - 100%