Chapter 13 monitoring 12-lead ecg – ZOLL Propaq MD Rev A User Manual

Page 151

Advertising
background image

9650-0805-01 Rev. A

Propaq MD Operator’s Guide

13-1

Chapter 13

Monitoring 12-Lead ECG

The Propaq MD 12-Lead input is Type CF defibrillator proof.

This chapter describes how to use the Propaq MD unit to perform 12-Lead ECG Monitoring.

The Propaq MD with the 12-lead ECG Monitoring option provides simultaneous acquisition
and storage of 12-lead information.

Warning!

12-Lead ECG monitoring is intended for the recording of 12-lead ECG signals
acquired from adult and pediatric patients in the supine, resting position -- always
ensure that the patient is kept motionless during 12-lead ECG signal acquisition. Use of
the device to acquire ECG signals from moving or shaking patients may produce noisy
signals that are difficult to interpret.

Excessive body hair or wet, sweaty skin may interfere with electrode adhesion. Remove
the hair and/or moisture from the area where the electrode is to be attached.

Remove ECG electrodes from their sealed package immediately prior to use. Using
previously opened or out-of- date electrodes may degrade the ECG signal quality.

Monitoring electrodes may become polarized during defibrillator discharge, causing
the ECG waveform to briefly go off screen. ZOLL Medical Corporation recommends
the use of high quality silver/silver chloride (Ag/AgCl) electrodes to minimize this
effect; the circuitry in the instrument returns the trace to the monitor display within a
few seconds.

Wait 15 seconds after defibrillator discharge before attempting a 12-lead acquisition.
Electrode polarization subsequent to defibrillator discharge may result in excessive
noise on the 12-lead ECG printout.

Advertising